If you have a thermometer with external calibration option (standard on ADJ models) you can recalibrate the thermometer as follows:
- Locate the 5/64 HEX set screw just under the head of the thermometer.
- Loosen the HEX set screw 1/2 to 1 turn.
- Immerse thermometer stem in minimum of two inches in an agitated liquid of known temperature for two to three minutes.
- Hold thermometer head and stem. Slowly turn thermometer head until pointer is at the temperature desired.
- Retighten HEX set screw.
NOTE: For accurate setting of the thermometer use a very accurate thermometer to verify the temperature of your liquid medium used to calibrate the thermometer.